4-channel Instrumentation Amplifier WGC-140A
Add to Wishlist1 in stock
Contact for a quote.
Request for Quotationช่องทางการสั่งซื้ออื่น ๆ
ติดต่อทาง Line
@santatech
ติดต่อทางอีเมล์
santatech@gmail.com
ติดต่อทางโทรศัพท์
02-1013454
Description
- Up to 4 units of 350 Ω strain-gage type transducers connected independently to respective amplifier circuits.
- Switchable indicated value, channel to channel or the totalAnalog voltage output of all channels and the total
- High/low limit comparator for the total value
- High-speed output suitable for control, I/O delay approx. 10 ms
- Sensor check and self-test functions
- Indication range-19999 to 19999 with decimal point anywhere
Features
The WGC-140A is a high-speed processing at 2 ms, a compact 4-channel Instrumentation Amplifier, which can calculate the physical quantities of load cells, etc.
It is suitable for measurement and control of the load applied to each column of press and the total load or for tension control by measurement of load balance and total load in various fields.
Specifications
Channels
Max. 4
Applicable Transducers
Strain-gage transducers
Compatible Bridge Resistance
350 to 1000 Ω
Measuring Range
±3 mV/V (Including tare)
Input LPF
150 Hz
Bridge Excitation
5 VDC
Digital Zero
Set at any point in the measuring range
Calibration
By inputting the value (By every channel)
By applying an actual load (By every channel)
Indicator
±19999
(Decimal point to be put anywhere, the same position is applied to all channels.)
Character height: 14 mm, red LED
Indicated value:
Measured value of a channel selected from 1 to 4
Total of measurements of channels 1 to 4
Calculation Speed
2 ms (500 times/s)
Nonlinearity
Within ±0.05% FS
Stability
Zero point: Within±0.5 μVRTI/°C
Sensitivity: Within±0.01%/°C
High/Low Limit Comparator
Setting points: 2 (High/low limits of total value)
Setting range: ±19999
Max. hysteresis width: 0 to 19999 settable
Output modes: Open collector (NPN)
Load capacity: 30 VDC, 20 mA (Resistive load)
Response speed: 10 ms or less
Smoothing Functions
Minimum scale:
Selectable from 1, 2, 5, 10, 20, 50, 100, 200, 500 or 1000 counts
Moving averaging functions:
Selectable from 2, 4, 8, 16, 32, 64, 128 or 256 times
Adding Functions
Setting range: ±19999
Original Value Measurement
±3 mV/V or more
Control Input
Input signals: 6
(4 calculation channel select commands, 1 each ZERO and CHECK commands)
Input modes: Non-voltage contact signal or open collector (NPN)
(It should enable application of 12 VDC and current flow of 5 mA.)
Control Output
Output signals: 7
(1 HEALTHY signal, 4 ABNORMAL signals and high/low limit signals)
Output modes: Open collector (NPN)
Load capacity: 30 VDC, 20 mA (Resistive load)
Analog Output
Output signals: 5 (Signals of 4 channels and the total)
Output voltage: ±10 V
Withstand voltage: 250 VAC for one minute
Resolution: 13 bits
Nonlinearity: Within ±0.1% FS
Conversion rate: 500 times/s
Setting values:
Indicated value with 0 V output (±19999)
Indicated value with 10 V output (±19999)
Check Functions
Self-test: Tests the program checksum and memory.
Transducer test:
Checks each channel for the bridge current, over-input and disconnection of transducer cable.
I/O Terminal Board
Transducer input: NDIS4102 (7 pins) connector plugs
Power connector, etc.: M3 screw terminal board
(Applicable crimp-style terminal V1.25-3 or the equivalent)
Data output terminal: Connector 57-40360 (DDK) or the equivalent
Operating Temperature
-10 to 55°C
Operating Humidity
20 to 85% (Non-condensing)
Power Supply
100 to 240 VAC, approx. 30 VA or less
Dimensions
96 W × 96 H ×139 D mm (Excluding protrusions)
Weight
Approx. 1.2 kg
Panel-cut Dimensions
92 × 92 mm
Standard Accessories
Instruction manual
Unit seal
Optional Accessories
AC power cables
P-23 (For 100 VAC)
P-28 (For 200 VAC)
SANTA TECHNOLOGY CO.,LTD. (Head Office)
35/218 Moo. 2, Leabkhlongsam Rd., Khlongsam, Khlongluang, Pathumthani 12120
Call 02-101-3454 ,089-006-2766
Fax 02-101-3453
Mail. santatech@gmail.com
Line : @santatech
Web : https://santatechnology.com/