Description
- Wide no-load zero adjustment range (±2 mV/V)
- High/low limit comparators
- Indicated value is output in a voltage range of 0 to 10 V or in a current range of 4 to 20 mA.
- Up to 4 transducers with 350 Ω bridge resistance connected in parallel.
The WGA-650B is an Instrumentation Amplifier which can display the physical quantity of an input sensor.
It is compact and moderate price.
Models
|
WGA-650B-0 M448: Power supply 100 to 127 VAC
WGA-650B-0 M449: Power supply 220 to 240 VAC
|
Specifications
Channels
|
1
|
Applicable Transducers
|
Strain-gage transducers (Bridge resistance of 350 Ω)
(Up to 4 transducers connectable in parallel)
|
Measuring Range
|
0 to 2.5 mV/V (5000 ×10-6 strain)
|
Bridge Excitation
|
10 VDC/2 VDC, switchable
|
No-load Zero Adjustment
|
±2 mV/V
Analog adjustment and digital adjustment in combination
|
Calibration
|
By inputting a numeric value
By applying an actual load
|
Display
|
-1999 to 19999 (Decimal point to be put anywhere.)
Character height 15.24 mm, red LED
|
Sampling Speed
|
4 times/s
|
Nonlinearity
|
Within ±(0.03% FS + 1 digit)
|
Stability
|
Zero point: ±0.5 μVRTI / ºC
Sensitivity: ±0.0025% / ºC
|
High/Low Limit Comparator
|
Setting points: 2 (High limit, low limit)
Setting range: -1999 to 19999
Contact output: Relay contact output (1 transfer circuit for each point)
Contact capacity: 250 VAC, 0.5 A (Resistive load)
|
Analog Filters
|
Cutoff frequencies: 1 Hz
|
Smoothing
|
Moving averaging times: 2, 4, 8, and 16
Min. scale: 1, 2, 5, 10, 20, 50, and 100
|
Zero Compensation
|
Digital zero compensation:
Possible with external voltage signals (10 to 30 VDC)
Automatic zero compensation:
Indication is made zero when the reading is in the setting range for 2 s or more.
|
Adding Functions
|
Setting range: -1999 to 19999
|
Analog (DA) Output
|
Voltage: 0 to 10 V (Load resistance 2 kΩ or more)
Desired scaling possible
Current: 4 to 20 mA (Load resistance 500 Ω or less)
Corresponds to voltage output of 0 to 10 V.
Withstand voltage between the output and case:
500 VAC for one minute
|
Check Functions
|
Transducer check
Self-check
|
Input Terminal Board
|
M3 screw board
(Applicable crimp-style terminals 1.25-3A and 1.25-3B or the equivalent)
|
Operating Temperature
|
-10 to 50°C
|
Operating Humidity
|
20 to 85% (Non-condensing)
|
Power Supply
|
100 to 127 VAC or 220 to 240 AC, 20 VA or less
|
Dimensions
|
96 W × 96 H ×139 D mm (Excluding protrusions)
|
Weight
|
Approx. 1.3 kg
|
Panel-cut Dimensions
|
92 × 92 mm
|
Standard Accessories
|
AC power cable P-23 (For 100 VAC)
Instruction manual
Unit seal
|
Optional Accessories
|
AC power cable P-28 (For 200 VAC)
Input cable U-33 (50 cm)
Input cable U-34 (1 m)
Input cable U-35 (2 m)
Input cable U-36 (5 m)
BCD output connector BCD-CONNE
(57-30360 (DDK) or the equivalent)
|
Models
|
WGA-650B-1 M448: Power supply 100 to 127 VAC
WGA-650B-1 M449: Power supply 220 to 240 VAC
|
Specifications
Channels
|
1
|
Applicable Transducers
|
Strain-gage transducers (Bridge resistance of 350 Ω)
(Up to 4 transducers connectable in parallel)
|
Measuring Range
|
0 to 2.5 mV/V (5000 ×10-6 strain)
|
Bridge Excitation
|
10 VDC/2 VDC, switchable
|
No-load Zero Adjustment
|
±2 mV/V
Analog adjustment and digital adjustment in combination
|
Calibration
|
By inputting a numeric value
By applying an actual load
|
Display
|
-1999 to 19999 (Decimal point to be put anywhere.)
Character height 15.24 mm, red LED
|
Sampling Speed
|
4 times/s
|
Nonlinearity
|
Within ±(0.03% FS + 1 digit)
|
Stability
|
Zero point: ±0.5 μVRTI / ºC
Sensitivity: ±0.0025% / ºC
|
High/Low Limit Comparator
|
Setting points: 2 (High limit, low limit)
Setting range: -1999 to 19999
Contact output: Relay contact output (1 transfer circuit for each point)
Contact capacity: 250 VAC, 0.5 A (Resistive load)
|
Analog Filters
|
Cutoff frequencies: 1 Hz
|
Smoothing
|
Moving averaging times: 2, 4, 8, and 16
Min. scale: 1, 2, 5, 10, 20, 50, and 100
|
Zero Compensation
|
Digital zero compensation:
Possible with external voltage signals (10 to 30 VDC)
Automatic zero compensation:
Indication is made zero when the reading is in the setting range for 2 s or more.
|
Adding Functions
|
Setting range: -1999 to 19999
|
Analog (DA) Output
|
Voltage: 0 to 10 V (Load resistance 2 kΩ or more)
Desired scaling possible
Current: 4 to 20 mA (Load resistance 500 Ω or less)
Corresponds to voltage output of 0 to 10 V.
Withstand voltage between the output and case:
500 VAC for one minute
|
BCD Output
|
Output mode: Isolated open collector (NPN)
BCD output logic selectable
Driving capacity: 30 VDC, 20 mA max
Output signals:
5-digit BCD value
Minus sign
OVER
Print command (EOC)
Connector: 57-40360 (DDK) or the equivalent
|
Check Functions
|
Transducer check
Self-check
|
Input Terminal Board
|
M3 screw board
(Applicable crimp-style terminals 1.25-3A and 1.25-3B or the equivalent)
|
Operating Temperature
|
-10 to 50°C
|
Operating Humidity
|
20 to 85% (Non-condensing)
|
Power Supply
|
100 to 127 VAC or 220 to 240 AC, 20 VA or less
|
Dimensions
|
96 W × 96 H ×139 D mm (Excluding protrusions)
|
Weight
|
Approx. 1.3 kg
|
Panel-cut Dimensions
|
92 × 92 mm
|
Standard Accessories
|
AC power cable P-23 (For 100 VAC)
Instruction manual
Unit seal
|
Optional Accessories
|
AC power cable P-28 (For 200 VAC)
Input cable U-33 (50 cm)
Input cable U-34 (1 m)
Input cable U-35 (2 m)
Input cable U-36 (5 m)
BCD output connector BCD-CONNE
(57-30360 (DDK) or the equivalent)
|
SANTA TECHNOLOGY CO.,LTD. (Head Office)
35/218 Moo. 2, Leabkhlongsam Rd., Khlongsam, Khlongluang, Pathumthani 12120
Call 02-101-3454 ,089-006-2766
Fax 02-101-3453
Mail. santatech@gmail.com
Line : @santatech
Web : https://santatechnology.com/