Compact Recording System EDX-10 Series
Add to Wishlist1 in stock
Contact for a quote.
Request for Quotationช่องทางการสั่งซื้ออื่น ๆ



Description
The EDX-10 Series is measuring instrument, which can measure strain, voltage, and temperature (Thermocouples) simply by connecting a PC via USB interface.
Up to 4 measuring units for 16 channels can be powered by USB interface, no separated power supply is required.
EDX-10B
Specifications
Interfaces
USB2.0 compliant
Connector configuration: USB standard B receptacle
Installed Measuring Units
Max. 4 (16 channels)
Sampling Frequencies
1 Hz to 20 k Hz (1 to 4 channels)
1 Hz to 10 k Hz (1 to 8 channels)
1 Hz to 5 k Hz (1 to 16 channels)
Operating Temperature
0 to 40ºC
Operating Humidity
20 to 90% (Non-condensing)
Power Supply
5 VDC by USB bus power or an optional accessory AC adapter
Notes:
1. When using USB bus-power, please connect directly to the USB port of the PC without using a USB hub.
2. When using AC adapter, the maximum connection units is 4 units in any
combination of EDX-11A, 12A, 13A, 14A and 15A.
Current Consumption
140 mA or less (5 VDC)
Weight
Approx. 170 g
Dimensions
84.0 W x 26.6 H x 84.0 D mm (Excluding protrusions)
Control Software
DCS-100A
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
USB cable N-38 (1 m)
Ground wire P-72 (5 m)
DVD (Dynamic data acquisition software DCS-100A)
Optional Accessories
AC adapter UN310-0515 (For U.S.A.: UN312-0520)
USB Ports
The number of measuring units at USB bus-power operation, refer to the following table.
【USB3.0】--------------------------------------------------------------------------------
EDX-11A units Number of EDX-12A/13A/14A/15A Connection units
0 Up to 4 units in any combination Max. 4
1 Until either 1 unit Max. 2
1 0 Max. 2
【USB2.0】--------------------------------------------------------------------------------
EDX-11A units Number of EDX-12A/13A/14A/15A Connection units
0 Up to 2 units in any combination Max. 2
1 0 Max. 1
EDX-14A
Specifications
Measuring Targets
Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.
Channels
4
Measuring Range
10 k, 50 k ×10-6 strain (2 steps)
Compatible Bridge Resistance
120 Ω to 1 kΩ
Bridge Excitation
1 VDC
Gage Factor
2.00 fixed
Range Accuracy
Within ±0.3%FS
Balance Adjustment
Within ±1/2 F.S. of setting range
Nonlinearity
Within ±0.1%FS
AD Converter
24 bits
Frequency Response
DC to 2 kHz
LPF
Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature
0 to 40ºC
Operating Humidity
20 to 90% (Non-condensing)
Input Connectors
D-sub 37-pin connector
Power Supply
5 VDC supplied by control unit
Current Consumption
140 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight
Approx. 150 g
Dimensions
84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
Strain input cable U-124 (30 cm)
Optional Accessories
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350
One-touch type input adapter UI-55A
Specifications
Measuring Targets
Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.
Voltage
Channels
4
Measuring Range
Strain-gage transducers, strain gages:
10 k, 50 k ×10-6 strain (2 steps)
Voltage: 10, 50 V
Compatible Bridge Resistance
120 Ω to 1 kΩ
Bridge Excitation
1 VDC
Gage Factor
2.00 fixed
Range Accuracy
Within ±0.3%FS
Balance Adjustment
Within ±1/2 F.S. of setting range
Nonlinearity
Within ±0.1%FS
AD Converter
24 bits
Frequency Response
DC to 2 kHz
LPF
Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature
0 to 40ºC
Operating Humidity
20 to 90% (Non-condensing)
Input Connectors
D-sub 37-pin connector
Power Supply
5 VDC supplied by control unit
Current Consumption
150 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight
Approx. 150 g
Dimensions
84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
Strain input cable U-124 (30 cm)
Conversion adapter FV-1A x 4
Optional Accessories
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350
One-touch type input adapter UI-55A
BNC input cable U-125 (30 cm)
EDX-12A
Specifications
Measuring Targets
Voltage
Channels
4 (Single end)
Measuring Range
10 V, 50 V (2 steps)
Range Accuracy
Each range within ±0.3%FS
Balance Adjustment
Within ±1/2 FS of setting range
Nonlinearity
Within ±0.1%FS
AD Converter
24 bits
Frequency Response
DC to 2 kHz
LPF
Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature
0 to 40ºC
Input Connectors
D-sub 37-pin connector
Power Supply
5 VDC supplied by control unit
Current Consumption
110 mA or less (5 VDC)
Weight
Approx. 150 g
Dimensions
84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
Input adapter UI-51A
Optional Accessories
BNC input cable U-125 (30 cm)
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
One-touch type input adapter UI-52A
EDX-13A
Specifications
Measuring Targets
Thermocouples
Channels
4
Input Connectors
Screw type terminal box
Applicable Thermocouples
K, T, J, N
(Resistance of thermocouple: 1 kΩ or less)
Range
K: -200.0 to 1370.0ºC
T: -200.0 to 400.0ºC
J: -200.0 to 1200.0ºC
N: -200.0 to 1300.0ºC
Resolution
0.1°C
Accuracy
[K]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1370.0°C: ± (0.1% of reading + 0.4ºC)
[T]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 400.0°C: ± (0.1% of reading + 0.4ºC)
[J]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1200.0°C: ± (0.1% of reading+0.4ºC)
[N]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1300.0°C: ± (0.1% of reading+0.4ºC)
Note: The measurement accuracy does not include the accuracy of the internal reference junction compensator and thermocouples.
Accuracy of internal reference junction compensator
±1.0°C
〔Input terminal temperature at equilibrium〕
〔Ambient temperature: 25 ±10°C〕
Mount the EDX-13A on the bottom when using it with measuring units other than the EDX-13A.
±2.0°C
〔Input terminal temperature in equilibrium〕
For temperatures other than those in the ambient temperature and operating temperature described above
Check Functions
Burnout check
AD Converter
24 bits
Sampling System
Scanning
Inside Samping Frequencies
Approx. 0.5 Hz, approx. 2.0 Hz
Operating Temperature
0 to 40℃
Operating Humidity
20 to 90% (Non-condensing)
Input Connector
Screw type terminal box
Power Supply
5 VDC supplied by control unit
Current Consumption
120 mA or less (5 VDC)
Weight
Approx. 130 g
Dimensions
84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
Terminal box 1 piece
Screwdriver 1 piece
EDX-11A
Specifications
Measuring Targets
Strain-gage transducers, strain gage
*Bridge boxes or input adapters are required for strain measurement.
Channels
4
Measuring Range
10 k, 50 k ×10-6 strain (2 steps)
Balance Adjustment
Within ±1/2 F.S. of setting range
Compatible Bridge Resistance
120 Ω to 1 kΩ
Bridge Excitation
2 VDC
Gage Factor
2.00 fixed
Range Accuracy
Each range within ±0.3%FS
Nonlinearity
Within ±0.1%FS
AD Converter
24 bits
Frequency Response
DC to 2 kHz
LPF
Cutoff frequencies : Lo (100 Hz), Hi (2 k Hz)
Transfer characteristic : 2nd order Butterworth
Operating Temperature
0 to 40ºC
Operating Humidity
20 to 90% (Non-condensing)
Input Connectors
D-sub 37-pin connector
Power Supply
5 VDC supplied by control unit
Current Consumption
180 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight
Approx. 150 g
Dimensions
84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance
Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories
Strain input cable U-124 (30 cm)
Optional Accessories
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A, UI-55A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350