Constant Current Amplifier Card CDA-44AS

SKU: CDA-44AS
Conditioner cards for EDX-100A EDX-200A and EDX-5000A. EDX-5000A does not support CAN card CAN-41A.

Description

The cards are for measuring strain, voltage, acceleration, and temperature. CAN data is supported. Isolated amplifier cards and carrier-wave amplifier cards also available for countermeasure against noise.

Specifications

Measuring Targets Strain gages (Full bridge system)
Strain-gage transducers
Voltage
Channels 4
Input Resistance Approx. 10 MΩ + 10 MΩ (Strain mode)
Approx. 1 MΩ (Voltage mode)
Input Format Balanced differential input (Strain mode)
Unbalanced input (Voltage mode)
IMRR 120 dB (When 500 ×10-6 strain range)
Frequency Response [DC coupling]
DC to 200 Hz, deviation: +1 dB, -3 dB

[DC cut (AC coupling)]
0.2 Hz (See the HPF)
Gage Factor 2.00 fixed (Strain mode)
Compatible Bridge Resistance 120 Ω
Bridge Excitation Approx. DC 16.7 mA (Constant current) when gage resistance 120 Ω connected

*If sensitivity or temperature resistance is in the transducer bridge excitation lines, then sensitivity and temperature characteristics are not corrected.
Cable Length 500 m (At cross section: 0.5 mm2)
Range Accuracy Within ±0.3% FS
Measuring Range 500, 1 k, 2 k, 5 k, 10 k, 20 k ×10-6 strain, OFF (Strain mode)
1, 2, 5, 10, 20, 50 V, OFF (Voltage mode)
Balance Adjustment Within ±2.4% (±12000 ×10-6 strain) (At strain measurement)
Within ±5 V (At voltage measurement)
ZERO Accuracy Within ±0.3% FS (Voltage OFF mode)
Nonlinearity Within ±0.1% FS
Calibration (CAL) Output at ±100% and ±50% of each range
Accuracy: Within ±0.3%FS
Monitor Output Accuracy: Within ±5 V ±0.5%
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 1, 3, 10, 30, 100 Hz and FLAT (6 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct.
HPF Cutoff frequencies : 0.2 Hz
Attenuation : Within (-6 ±1) dB/oct.
AD Converter 16 bits
TEDS Reads information from TEDS-installed sensors.
Isolation Between input and case (output): Withstand voltage 500 VDC,1 min.
Between channels: Withstand voltage 500 VDC,1 min.
Standard Accessories Conversion adapter FV-2A × 4
Optional Accessories Monitor output cable U-64
Note If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement.
But the remote-sensing function will be ineffective.

SANTA TECHNOLOGY CO.,LTD. (Head Office)
35/218 Moo. 2, Leabkhlongsam Rd., Khlongsam, Khlongluang, Pathumthani 12120
Call 02-101-3454 ,089-006-2766
Fax 02-101-3453
Mail. santatech@gmail.com
Line : @santatech
Web :
https://santatechnology.com/

ติดต่อสอบถาม คลิก
อีเมล์
โทรศัพท์
แชทผ่าน LINE
แชทผ่าน Facebook
ดูแผนที่

please wait.....