Hanna pH/mV Meter Model HI2002-Edge
Add to Wishlist1 in stock
Contact for a quote.
Request for Quotationช่องทางการสั่งซื้ออื่น ๆ
ติดต่อทาง Line
@santatech
ติดต่อทางอีเมล์
santatech@gmail.com
ติดต่อทางโทรศัพท์
02-1013454
Description
Hanna pH/mV Meter Model HI2002-Edge
edge’s groundbreaking design is the culmination of Hanna’s vision, design capabilities, integrated production and world class R&D. edge is a single, easy to use meter that can measure pH and ORP.
Details
Digital pH Electrodes
edge pH electrodes are highly advanced, featuring a built-in microchip that stores calibration information that is automatically retrieved by edge once the electrode is plugged in. With a digital electrode it is possible to change electrodes without having to recalibrate since the calibration information is stored in the pH electrode and not the meter. This allows flexibility for the user that measures pH with more than one electrode.
CAL Check
Hanna’s exclusive CAL Check feature analyses the pH electrode response in the pH buffers during the calibration process to alert the user of potential problems such as contaminated buffer and clean electrode. After calibration, indicators for probe condition and response time are displayed on the measurement screen. The probe condition indicator is based on offset and slope characteristics of the pH electrode.
Sensor Check
-When used with Hanna’s electrodes equipped with a matching pin, edge constantly checks the impedance of the pH measuring electrode to notify you in real time in the event of glass breakage. During calibration, Sensor Check checks the state of the junction. The reference junction is also evaluated and reported on the display.
Wall mount
-Using the wall mount cradle (included), edge can be placed on a wall, leaving zero footprint on the benchtop space. The cradle has a built-in connector to power edge and charge its batteries. edge's zero footprint is designed to save you valuable benchtop space.
Cradle and Electrode Holder
-edge is equipped with a benchtop cradle with an adjustable swivel electrode holder to charge and hold the edge securely in place at the optimum viewing angle.
Edge thin pH meter handheld
-edge® is incredibly thin and weighs only 8.8 oz., has an 8 hour battery life, and dual USB inputs.
Two USB Ports
-edge includes one standard USB for exporting data with a flash drive and a second micro USB port for you to connect to a computer for file export and for charging your edge when the cradle is not available.
Data Logging
-edge allows you to store up to 1000 log records of data. Data sets include readings, GLP data, date and time.
Two Operating Modes
-edge can be used in Extended or Basic Operating Modes. Extended Mode enables all features available while the Basic Mode reduces the amount of features to a very simple operation. For example, in extended mode edge offers 5 point calibration with 2 custom buffers while in basic mode calibration can only be performed to 3 points with 5 preprogrammed buffers.
pH Specifications
- - pH Range -2.000 to 16.000 pH (Standard Mode), -2.00 to 16.00 pH (Basic Mode)
- - pH Resolution 0.001 pH, 0.01 pH
- - pH Accuracy (@25ºC/77ºF) ±0.01 pH, ±0.002 pH
- - pH Calibration 5 points (Standard mode) 1.68, 4.01 (3.00†), 6.86, 7.01, 9.18, 10.01, 12.45, and two custom buffers; 3 points (Basic mode) 4.01; 6.86; 7.01; 9.18; 10.01
- - pH Temperature Compensation ATC: -5.0 to 100.0ºC; 23.0 to 212.0°F*
- - mV Range ±1000.0 mV; ±2000.0 mV
- - mV Resolution 0.1 mV
- - mV Accuracy ±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)
- - Relative mV Calibration Single point calibration
Temperature Specifications
- Temperature Range -20.0 to 120.0 ºC, -4.0 to 248.0 °F
- Temperature Resolution 0.1 °C, 0.1 °F
- Temperature Accuracy ±0.5 °C, ±0.9 °F
- °C/°F Yes
Additional Specifications
pH Electrode Diagnostics Glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time
GLP Yes
Logging up to 1000*** records organized in: Manual log-on-demand (Max. 200 logs), Manual log-on-stability (Max. 200 logs), Interval logging *** (Max. 600 samples; 100 lots)
Connectivity 1 micro USB port for charging and PC connectivity, 1 USB port for storage
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Battery Type/Life Built-in rechargeable battery with up to 8 hours of continuous use
Power Supply 5 VDC adapter (included)
Dimensions 202 x 140 x 12.7mm (8” x 5.5” x 0.5”)
Weight 250 g (8.82 oz)
Ordering Information
edge® HI2002 includes benchtop docking station with electrode holder, wall mount cradle, USB cable, HI11310 glass body refillable pH electrode with temperature sensor, 2 sachets each of pH 4, 7, 10, cleaning solution, 5 VDC power adapter and instruction manual.
Notes
* temperature limits will be reduced to actual probe/sensor limits
*** For standard mode only (logging)